대동맥 인조혈관 원형 스테이플 문합에서의 누출 위험 분석 : 유한요소 시뮬레이션 기반 연구
Leakage Risk Analysis in Aortic Graft Circular Stapler Anastomosis: A Finite Element Simulation Study

초록

Aortic aneurysms, caused by abnormal expansion of the aortic wall, pose critical risks upon rupture. Cir- cular stapler anastomosis offers a faster and safer alternative to suturing. This study evaluates leakage risk and fix- ation strength in aortic graft circular stapler anastomosis using finite element analysis (FEA). A simplified FEA model was developed and validated using existing leakage pressure data. Key parameters, including inter-row distance, tri- ple-row structure and triple-row spacing were analyzed. The Factor of Safety (FoS) was defined to quantitatively com- pare leakage risks under varying conditions. The inter-row distance critically influenced leakage risk, with wider gaps increasing leakage likelihood. Conversely, the triple-row structure significantly enhanced fixation strength, reducing leakage risk. Internal spacing within the triple-row exhibited minimal sensitivity. This study establishes a design framework for safer and more effective aortic graft circular stapler anastomosis. Findings provide foundational data for optimizing design and assessing leakage risk.

키워드

Aortic aneurysmCircular staplerLeakage riskFinite element analysisFactor of safety
제목
대동맥 인조혈관 원형 스테이플 문합에서의 누출 위험 분석 : 유한요소 시뮬레이션 기반 연구
제목 (타언어)
Leakage Risk Analysis in Aortic Graft Circular Stapler Anastomosis: A Finite Element Simulation Study
저자
서명재권지연김성민
DOI
10.9718/JBER.2025.46.4.366
발행일
2025-08
유형
Y
저널명
의공학회지
46
4
페이지
366 ~ 375