Compound Acoustic Radiation Force Impulse Imaging of Bovine Eye by Using Phase-Inverted Ultrasound Transducer
- Kim, Gil Su;
- Moon, Hak Hyun;
- Lee, Hee Su;
- Jeong, Jong Seob
WEB OF SCIENCE
4SCOPUS
3초록
In general, it is difficult to visualize internal ocular structure and detect a lesion such as a cataract or glaucoma using the current ultrasound brightness-mode (B-mode) imaging. This is because the internal structure of the eye is rich in moisture, resulting in a lack of contrast between tissues in the B-mode image, and the penetration depth is low due to the attenuation of the ultrasound wave. In this study, the entire internal ocular structure of a bovine eye was visualized in an ex vivo environment using the compound acoustic radiation force impulse (CARFI) imaging scheme based on the phase-inverted ultrasound transducer (PIUT). In the proposed method, the aperture of the PIUT is divided into four sections, and the PIUT is driven by the out-of-phase input signal capable of generating split-focusing at the same time. Subsequently, the compound imaging technique was employed to increase signal-to-noise ratio (SNR) and to reduce displacement error. The experimental results demonstrated that the proposed technique could provide an acoustic radiation force impulse (ARFI) image of the bovine eye with a broader depth-of-field (DOF) and about 80% increased SNR compared to the conventional ARFI image obtained using the in-phase input signal. Therefore, the proposed technique can be one of the useful techniques capable of providing the image of the entire ocular structure to diagnose various eye diseases.
키워드
- 제목
- Compound Acoustic Radiation Force Impulse Imaging of Bovine Eye by Using Phase-Inverted Ultrasound Transducer
- 저자
- Kim, Gil Su; Moon, Hak Hyun; Lee, Hee Su; Jeong, Jong Seob
- 발행일
- 2024-05
- 유형
- Article
- 저널명
- Sensors
- 권
- 24
- 호
- 9
- 페이지
- 1 ~ 15
- 언어
- ENG
- 출판사
- Multidisciplinary Digital Publishing Institute (MDPI)
- 발행국가
- 스위스
- 분량
- 15 페이지
- ISSN
- E 1424-8220
P 1424-8220