Detection of HIV-1 antigen by quartz crystal microbalance using gold nanoparticles

Citations

WEB OF SCIENCE

37
Citations

SCOPUS

40

초록

In the current exploratory study, a quartz crystal microbalance (QCM) technique was employed to detect HIV-1 antigen at a very low concentration using gold nanoparticles (Au NPs) as a signal enhancer. Colloidal Au NPs were synthesized using a reduction method followed by streptavidin functionalization. HIV-1 antigen was captured by a polyclonal antibody on the QCM surface that had been previously treated with 11-mercaptoundecanoicacid (MUA) and streptavidin. The Au NPs with a high mass amplified the detection signal, making HIV-1 antigen detection possible even in the window period of HIV infection. In addition, the effect of Au NP size was also investigated to elucidate the optimal size of the enhancer. The signal performance is discussed in terms of mass effect and conjugate efficiency. (C) 2016 Elsevier B.V. All rights reserved.

키워드

HIV-1 antigen; Quartz crystal microbalance; Gold nanoparticles; Early detection; Signal amplification; STREPTAVIDIN; PERFORMANCE
제목
Detection of HIV-1 antigen by quartz crystal microbalance using gold nanoparticles
저자
Tan Nhiem Ly; Park, Sangkwon; Park, Sang Joon
DOI
10.1016/j.snb.2016.06.112
발행일
2016-12
유형
Article
저널명
Sensors and Actuators, B: Chemical
권
237
페이지
452 ~ 458