상세 보기
초록
We report our setup and measurement results of the quality factor Q of superconducting microwave resonators at mK temperature in a dilution refrigerator. The power dependence of the superconducting resonators' Q was measured down to the single photon level. Measurements were performed on two types of microwave resonators. First, a half-wavelength coplanar waveguide (CPW) resonator made of niobium film on an oxidized high-resistivity silicon substrate was measured. Then a rectangular box resonator made of machined superconducting aluminum alloy was measured. While the CPW chip resonator's Q value showed noticeable decrease at low microwave power, the box resonator's Q showed little power dependence. This setup is intended to perform characterization of various microwave resonators, which will serve as key components in superconducting qubit in circuit QED architecture. © 2014 IEEE.
키워드
- 제목
- Power-dependence measurement of the quality factor of superconducting microwave resonators at mK temperature
- 저자
- Kim, S.H.; Ha, D.-G.; Park, G.Y.; Choi, J.; Park, J.H.; Song, W.; Lee, S.G.; Im, H.; Kim, H.; Chong, Y.
- 발행일
- 2014-09-12
- 유형
- Conference Paper
- 저널명
- CPEM Digest (Conference on Precision Electromagnetic Measurements)
- 페이지
- 174 ~ 175