Dual-Frequency Ultrasound Transducer Using Inversion Layer Technique for Therapeutic Ultrasound Surgery

Citations

WEB OF SCIENCE

18
Citations

SCOPUS

18

초록

High intensity focused ultrasound surgery still requires relatively a long surgical time due to small size of coagulative necrosis per each sonication. In order to increase a region of coagulative necrosis, dual-frequency ultrasound scheme has been proposed and showed notable results. In this paper, a novel dual-frequency ultrasound transducer based on the inversion layer technique is presented. Finite element analysis simulation was conducted, and the prototype transducer was fabricated to demonstrate the performance of the proposed method. The optimized thickness ratio between the inverted and non-inverted piezoelectric layers was chosen as 0.3:0.7. The proposed transducer with the chosen ratio generated the fundamental (at 1.2 MHz) frequency and the second harmonic frequency (at 2.4 MHz) ultrasound at the same time. The performance of the proposed transducer was numerically and experimentally verified through electrical impedance, pulse-echo response, and beam profile tests.

키워드

Ultrasound transducer; high intensity focused ultrasound; dual-frequency ultrasound; inversion layer; INTENSITY FOCUSED ULTRASOUND; LINBO3 SINGLE-CRYSTAL; ACOUSTIC CAVITATION; TISSUE; LIVER; HIFU; EXPOSURE; PLATE; ENHANCEMENT; ABLATION
제목
Dual-Frequency Ultrasound Transducer Using Inversion Layer Technique for Therapeutic Ultrasound Surgery
저자
Park, Chan Yuk; Kwon, Da Sol; Sung, Jin Ho; Jeong, Jong Seob
DOI
10.1109/JSEN.2017.2752229
발행일
2017-11-01
유형
Article
저널명
IEEE Sensors Journal
권
17
호
21
페이지
6859 ~ 6866