Improvement of dual-glucose sensor specificity for prosthetic vascular grafts based on a calibration scheme

  • Jang, Heedon; 
  • Kim, Seongmun; 
  • Ma, Hyunggun; 
  • Patel, Ramesh; 
  • Yang, Seungboo; 
  • ... Han, Ki Jin; 
  • 외 3명
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

Glucose monitoring is an important clinical procedure, especially for dialysis patients who need consistent monitoring of their glucose levels. Currently, the most extensively used method for glucose monitoring involves pricking the finger and sampling a small amount of blood. Given that this procedure is inconvenient and can cause pain and potential infection, there is demand for the development of alternative glucose sensing methods. This study introduces a methodology for improved glucose sensor specificity based on a calibration scheme. One microwave and one capacitive glucose sensor were designed and placed on a prosthetic vascular graft. Each sensor yielded a finite variation in the measured glucose concentrations based on its capacity to sense permittivity changes in aqueous D-glucose solutions. However, as blood components other than glucose-such as proteins, erythrocytes and haemoglobin-may affect the measurements, the authors also introduced a calibration scheme to adjust and calibrate each measurement to ensure accuracy. The measurement data yielded a maximum error of <7.33%. Based on these outcomes, the specificity of glucose monitoring in prosthetic vascular grafts is validated.

키워드

patient monitoring; sugar; permittivity; cellular biophysics; prosthetics; biochemistry; blood; proteins; chemical sensors; biomedical measurement; calibration; molecular biophysics; calibration scheme; capacitive glucose sensor; prosthetic vascular graft; measured glucose concentrations; glucose monitoring; dual-glucose sensor specificity; clinical procedure; glucose levels; alternative glucose sensing methods; dialysis patients; microwave glucose sensor; permittivity changes; aqueous D-glucose solutions; blood components; proteins; erythrocytes; haemoglobin; MICROWAVE; SYSTEM
제목
Improvement of dual-glucose sensor specificity for prosthetic vascular grafts based on a calibration scheme
저자
Jang, Heedon; Kim, Seongmun; Ma, Hyunggun; Patel, Ramesh; Yang, Seungboo; Jeong, Jiyun; Seo, Jongmo; Han, Ki Jin; Bien, Franklin
DOI
10.1049/iet-map.2019.0617
발행일
2020-09-09
유형
Article
저널명
IET Microwaves, Antennas and Propagation
권
14
호
11
페이지
1222 ~ 1228