Chronic wounds necessitate periodic treatment and management due to their potential for serious complications. Recently,ultrasonic mist therapy has been introduced to treat chronic wounds efficiently. This therapy requires a noncontact sprayingmethod to prevent side effects such as bacterial infections and pain. Therefore, research is needed on a spray nozzle tip thatcan effectively transmit ultrasonic energy to the wound target with misted cleaning solution mobility in a specific direction andat an appropriate speed. The performance of the nozzle tip is greatly affected by the flow characteristics inside it. Computational fluid dynamics (CFD) is a powerful tool to analyze these characteristics in detail. The behavior of the mist wasanalyzed in a simulation based on discrete phase model methodology in an unsteady state. Valid design parameters enablingnoncontact cleaning were determined by setting the design parameters of the nozzle tip`s internal flow path and measuring thespraying speed of the mist using CFD analysis. Through the simulation results, information on the sprayed skin surface andspray characteristics are measured. Lastly, we present a nozzle tip design guide optimized for ultrasonic mist therapy.
키워드
- 제목
- 비접촉식 의료용 초음파미스트 분무를 위한 노즐 팁 성능 개선
- 제목 (타언어)
- Improvement of Nozzle Tip Performance for Noncontact Medical Ultrasonic Mist Spraying
- 저자
- 정승혁; 전종혁; 원지영; 김성민; 임홍석
- 발행일
- 2024-06
- 저널명
- 한국정밀공학회지
- 권
- 41
- 호
- 6
- 페이지
- 489 ~ 496