Quantitative analysis of PACAP-38 and amyloid beta composition using a dual reaction in reduced graphene oxide biosensor to verify the neuroprotective role of PACAP-38open access

Authors
Lee, SeungjunSong, Hyo RyeongPark, KeunwanLee, MinhyeongLee, A-HyeonRoh, Seung MinPark, YejinKim, Hye JinKwon, YoungeunYang, Seung-HoonKim, Jinsik
Issue Date
Jun-2022
Publisher
Elsevier BV
Keywords
PACAP-38; Amyloid beta; Reduced graphene oxide biosensor; Charged biomarker; Neuroprotection
Citation
Sensors and Actuators B: Chemical, v.361, pp 1 - 10
Pages
10
Indexed
SCIE
SCOPUS
Journal Title
Sensors and Actuators B: Chemical
Volume
361
Start Page
1
End Page
10
URI
https://scholarworks.dongguk.edu/handle/sw.dongguk/2951
DOI
10.1016/j.snb.2022.131727
ISSN
0925-4005
1873-3077
Abstract
Here, we developed a sophisticated reduced graphene oxide (rGO) sensor capable of monitoring the progress of Alzheimer's disease by quantifying the concentration of amyloid beta (A beta) and pituitary adenylate cyclase activating polypeptide 38 (PACAP-38). Western blot analysis following induction of apoptosis in mouse hippocampal neurons with TNF-alpha demonstrated that PACAP-38/A beta binding inhibited neural cell apoptosis. To verify the direct binding of PACAP-38 to A beta, a simulated docking model was generated, and binding affinities were quantified using surface plasmon resonance (SPR). After PACAP-38-A beta binding was confirmed, the properties of each protein were evaluated by the rGO sensor based on differences in their charge to determine the PACAP-38: A beta ratio. Based on the observed changes in resistance obtained by measuring each protein monomer, expected resistance change values were derived for each binding reaction and subsequent accumulation of PACAP-38 and A beta. Subsequently, rGO sensor measured changes in resistance following binding and accumulation of each protein. The resistance change values were compared with the expected values to verify the accuracy of the sensor for quantifying accumulated proteins. The accuracy of the sensor was determined based on deviation of the sensor output within its dynamic range. The significance was expressed based on the equation for significance values and was confirmed to be as high as 96% between expected and measured values. Collectively, our results verify the neuroprotective role of PACAP-38 and proposes rGO sensor as a potential tool for early diagnosis of as well as monitoring the progression of Alzheimer's disease.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Life Science and Biotechnology > Department of Biomedical Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kwon, Young Eun photo

Kwon, Young Eun
College of Life Science and Biotechnology (Department of Biomedical Engineering)
Read more

Altmetrics

Total Views & Downloads

BROWSE