Detailed Information

Cited 5 time in webofscience Cited 6 time in scopus
Metadata Downloads

Stress Distribution due to Loading on Premolar Teeth Implant: A Three Dimensional Finite Element Analysis

Full metadata record
DC Field Value Language
dc.contributor.authorHisam, M. J.-
dc.contributor.authorLim, J. Y.-
dc.contributor.authorKurniawan, D.-
dc.contributor.authorNor, F. M.-
dc.date.accessioned2024-08-08T12:01:24Z-
dc.date.available2024-08-08T12:01:24Z-
dc.date.issued2015-
dc.identifier.issn2351-9789-
dc.identifier.urihttps://scholarworks.dongguk.edu/handle/sw.dongguk/21984-
dc.description.abstractThis study investigates, with the use of three dimensional finite element analysis, the stress that occurred on peri-implant bone during loading on a premolar implant. The bone was assumed to be transversely isotropic. The refined method is implemented in a case study, in which osseointegrated (25-100%) peri-implant premolar bone types II, III and IV with an implant attached is loaded with a 100 MPa load. The results showed that the less dense bone must support higher stress and strain, especially at the cortical region. Higher degree of osseointegration caused higher stress but lower strain. Both the bone type and the osseointegration condition significantly affect the stress-strain relation. For minimum stress and strain, denser and more osseointegrated peri-implant bone is desirable. (C) 2015 Published by Elsevier B.V.-
dc.format.extent6-
dc.language영어-
dc.language.isoENG-
dc.publisherELSEVIER SCIENCE BV-
dc.titleStress Distribution due to Loading on Premolar Teeth Implant: A Three Dimensional Finite Element Analysis-
dc.typeArticle-
dc.publisher.location네델란드-
dc.identifier.doi10.1016/j.promfg.2015.07.038-
dc.identifier.scopusid2-s2.0-85013130700-
dc.identifier.wosid000380490700038-
dc.identifier.bibliographicCitation2ND INTERNATIONAL MATERIALS, INDUSTRIAL, AND MANUFACTURING ENGINEERING CONFERENCE, MIMEC2015, v.2, pp 218 - 223-
dc.citation.title2ND INTERNATIONAL MATERIALS, INDUSTRIAL, AND MANUFACTURING ENGINEERING CONFERENCE, MIMEC2015-
dc.citation.volume2-
dc.citation.startPage218-
dc.citation.endPage223-
dc.type.docTypeProceedings Paper-
dc.description.isOpenAccessY-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaEngineering-
dc.relation.journalWebOfScienceCategoryEngineering, Manufacturing-
dc.relation.journalWebOfScienceCategoryEngineering, Mechanical-
dc.subject.keywordPlusBONE-
dc.subject.keywordAuthorFinite element analysis-
dc.subject.keywordAuthorPremolar-
dc.subject.keywordAuthorDental implant-
dc.subject.keywordAuthorStress-
dc.subject.keywordAuthorStrain-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Engineering > Department of Mechanical, Robotics and Energy Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lim, Joong Yeon photo

Lim, Joong Yeon
College of Engineering (Department of Mechanical, Robotics and Energy Engineering)
Read more

Altmetrics

Total Views & Downloads

BROWSE