Cited 7 time in
Computational Wear Prediction of TKR with Flatback Deformity during Gait
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | Lee, Hye Kyeong | - |
| dc.contributor.author | Kim, Sung Min | - |
| dc.contributor.author | Lim, Hong Seok | - |
| dc.date.accessioned | 2024-08-08T10:02:07Z | - |
| dc.date.available | 2024-08-08T10:02:07Z | - |
| dc.date.issued | 2022-04 | - |
| dc.identifier.issn | 2076-3417 | - |
| dc.identifier.issn | 2076-3417 | - |
| dc.identifier.uri | https://scholarworks.dongguk.edu/handle/sw.dongguk/21372 | - |
| dc.description.abstract | Loss of lumbar lordosis in flatback patients leads to changes in the walking mechanism like knee flexion. Such variations in flatback patients are predicted to alter the characteristics of total knee replacement (TKR) contact, so their TKR will show different wear characteristics with a normal gait. However, the relevant study is limited to predicting the wear depth of TKR for normal gait mechanisms or collecting and analyzing kinematic data on flatback gait mechanisms. The objective of this study was to compare wear in TKR of flatback patients with people without flatback syndrome. The main difference between the normal gait mechanism and the flat back gait mechanism is the knee flexion remain section and the tendency to change the vertical force acting on the knee. Thus, in this paper, A finite element-based computational wear simulation for the gait cycle using kinematic data for normal gait and flat gait were performed, and substituting the derived contact pressure and slip distance into the Archard formula, a proven wear model, wear depth was predicted. The FE analysis results show that the wear volume in flatback patients is greater. The results obtained can provide guidance on the TKR design to minimize wear on the knee implant for flatback patients. | - |
| dc.format.extent | 15 | - |
| dc.language | 영어 | - |
| dc.language.iso | ENG | - |
| dc.publisher | MDPI | - |
| dc.title | Computational Wear Prediction of TKR with Flatback Deformity during Gait | - |
| dc.type | Article | - |
| dc.publisher.location | 스위스 | - |
| dc.identifier.doi | 10.3390/app12073698 | - |
| dc.identifier.scopusid | 2-s2.0-85128546373 | - |
| dc.identifier.wosid | 000781910600001 | - |
| dc.identifier.bibliographicCitation | Applied Sciences, v.12, no.7, pp 1 - 15 | - |
| dc.citation.title | Applied Sciences | - |
| dc.citation.volume | 12 | - |
| dc.citation.number | 7 | - |
| dc.citation.startPage | 1 | - |
| dc.citation.endPage | 15 | - |
| dc.type.docType | Article | - |
| dc.description.isOpenAccess | Y | - |
| dc.description.journalRegisteredClass | scie | - |
| dc.description.journalRegisteredClass | scopus | - |
| dc.relation.journalResearchArea | Chemistry | - |
| dc.relation.journalResearchArea | Engineering | - |
| dc.relation.journalResearchArea | Materials Science | - |
| dc.relation.journalResearchArea | Physics | - |
| dc.relation.journalWebOfScienceCategory | Chemistry, Multidisciplinary | - |
| dc.relation.journalWebOfScienceCategory | Engineering, Multidisciplinary | - |
| dc.relation.journalWebOfScienceCategory | Materials Science, Multidisciplinary | - |
| dc.relation.journalWebOfScienceCategory | Physics, Applied | - |
| dc.subject.keywordPlus | TOTAL KNEE REPLACEMENT | - |
| dc.subject.keywordPlus | SAGITTAL PLANE | - |
| dc.subject.keywordPlus | MANAGEMENT | - |
| dc.subject.keywordAuthor | knee implant | - |
| dc.subject.keywordAuthor | wear prediction | - |
| dc.subject.keywordAuthor | wear simulation | - |
| dc.subject.keywordAuthor | finite element analysis | - |
| dc.subject.keywordAuthor | flatback deformity | - |
| dc.subject.keywordAuthor | TKR design | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
30, Pildong-ro 1-gil, Jung-gu, Seoul, 04620, Republic of Korea+82-2-2260-3114
Copyright(c) 2023 DONGGUK UNIVERSITY. ALL RIGHTS RESERVED.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.