Cited 0 time in
Comparison of the GlideRite to the conventional malleable stylet for endotracheal intubation by the Macintosh laryngoscope: a simulation study using manikins
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | 공용택 | - |
| dc.contributor.author | 이현정 | - |
| dc.contributor.author | 나지웅 | - |
| dc.contributor.author | 신동혁 | - |
| dc.contributor.author | 한상국 | - |
| dc.contributor.author | 이정훈 | - |
| dc.contributor.author | 최필조 | - |
| dc.date.accessioned | 2024-08-08T03:30:58Z | - |
| dc.date.available | 2024-08-08T03:30:58Z | - |
| dc.date.issued | 2016-03 | - |
| dc.identifier.uri | https://scholarworks.dongguk.edu/handle/sw.dongguk/17042 | - |
| dc.description.abstract | Objective To compare the effectiveness of the GlideRite stylet with the conventional malleable stylet (CMS) in endotracheal intubation (ETI) by the Macintosh laryngoscope. Methods This study is a randomized, crossover, simulation study. Participants performed ETI using both the GlideRite stylet and the CMS in a normal airway model and a tongue edema model (simulated difficult airway resulting in lower percentage of glottic opening [POGO]). Results In both the normal and tongue edema models, all 36 participants successfully performed ETI with the two stylets on the first attempt. In the normal airway model, there was no difference in time required for ETI (TETI) or in ease of handling between the two stylets. In the tongue edema model, the TETI using the CMS increased as the POGO score decreased (POGO score was negatively correlated with TETI for the CMS, Spearman’s rho=-0.518, P=0.001); this difference was not seen with the GlideRite (rho=-0.208, P=0.224). The TETI was shorter with the GlideRite than with the CMS, however, this difference was not statistically significant (15.1 vs. 18.8 seconds, P=0.385). Ease of handling was superior with the GlideRite compared with the CMS (P=0.006). Conclusion Performance of the GlideRite and the CMS were not different in the normal airway model. However, in the simulated difficult airway model with a low POGO score, the GlideRite performed better than the CMS for direct laryngoscopic intubation. | - |
| dc.format.extent | 7 | - |
| dc.language | 영어 | - |
| dc.language.iso | ENG | - |
| dc.publisher | 대한응급의학회 | - |
| dc.title | Comparison of the GlideRite to the conventional malleable stylet for endotracheal intubation by the Macintosh laryngoscope: a simulation study using manikins | - |
| dc.title.alternative | Comparison of the GlideRite to the conventional malleable stylet for endotracheal intubation by the Macintosh laryngoscope: a simulation study using manikins | - |
| dc.type | Article | - |
| dc.identifier.doi | 10.15441/ceem.15.038 | - |
| dc.identifier.bibliographicCitation | Clinical and Experimental Emergency Medicine, v.3, no.1, pp 9 - 15 | - |
| dc.citation.title | Clinical and Experimental Emergency Medicine | - |
| dc.citation.volume | 3 | - |
| dc.citation.number | 1 | - |
| dc.citation.startPage | 9 | - |
| dc.citation.endPage | 15 | - |
| dc.identifier.kciid | ART002132693 | - |
| dc.description.isOpenAccess | N | - |
| dc.subject.keywordAuthor | Intubation | - |
| dc.subject.keywordAuthor | intratracheal | - |
| dc.subject.keywordAuthor | Instrumentation | - |
| dc.subject.keywordAuthor | Manikins | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
30, Pildong-ro 1-gil, Jung-gu, Seoul, 04620, Republic of Korea+82-2-2260-3114
Copyright(c) 2023 DONGGUK UNIVERSITY. ALL RIGHTS RESERVED.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.