Detailed Information

Cited 20 time in webofscience Cited 20 time in scopus
Metadata Downloads

Design and fabrication of ultrasound linear array transducer based on polarization inversion technique

Authors
Park, Chan YukSung, Jin HoJeong, Jong Seob
Issue Date
1-Sep-2018
Publisher
ELSEVIER SCIENCE SA
Keywords
Ultrasound array transducer; Polarization inversion technique (PIT); Finite element analysis (FEA) simulation; Broad bandwidth
Citation
SENSORS AND ACTUATORS A-PHYSICAL, v.280, pp 484 - 494
Pages
11
Indexed
SCI
SCIE
SCOPUS
Journal Title
SENSORS AND ACTUATORS A-PHYSICAL
Volume
280
Start Page
484
End Page
494
URI
https://scholarworks.dongguk.edu/handle/sw.dongguk/16974
DOI
10.1016/j.sna.2018.08.008
ISSN
0924-4247
1873-3069
Abstract
It has been well known that fabrication of the ultrasound array transducer is very difficult mainly due to the thin piezoelectric layer. Since the center frequency of the ultrasound array transducer is inversely proportional to the thickness of the piezoelectric layer, as the center frequency is increased, fabrication of the array transducer becomes more difficult. To solve this problem, in this paper, the polarization inversion technique (PIT) was employed to increase the center frequency of the array transducer without making the piezoelectric layer thin. Additionally, the PIT can increase the frequency bandwidth of the transducer by using its multiple-resonance feature. According to the finite element analysis (FEA) simulation, when the total thickness of the piezoelectric layer was fixed at 311 mu m, the conventional transducer had a center frequency of 4.5 MHz and -6 dB bandwidth of 71.4%, while the proposed transducer with PIT had a center frequency of 8.1 MHz and -6 dB bandwidth of 83.2%. In the case of conventional transducer with about 8 MHz center frequency, the thickness of the piezoelectric layer was 155 mu m, and -6 dB bandwidth was 72.3%. Subsequently, we fabricated a prototype transducer with PIT and verified that the center frequency and -6 dB bandwidth were 7.8 MHz and 81.3%, respectively. Therefore, the PIT can be a useful method for manufacturing the ultrasound array transducer with high frequency and broad bandwidth. (C) 2018 Elsevier B.V. All rights reserved.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Life Science and Biotechnology > Department of Biomedical Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Jeong, Jong Seob photo

Jeong, Jong Seob
College of Life Science and Biotechnology (Department of Biomedical Engineering)
Read more